The American Psychiatric Association (APA) has updated its Privacy Policy and Terms of Use, including with new information specifically addressed to individuals in the European Economic Area. As described in the Privacy Policy and Terms of Use, this website utilizes cookies, including for the purpose of offering an optimal online experience and services tailored to your preferences.

Please read the entire Privacy Policy and Terms of Use. By closing this message, browsing this website, continuing the navigation, or otherwise continuing to use the APA's websites, you confirm that you understand and accept the terms of the Privacy Policy and Terms of Use, including the utilization of cookies.

×
Published Online:

To the Editor: Drug screening through urinalysis is a widely accepted method for rapid detection of potential drug use at a relatively low cost. On the other hand, interferences are not that unusual in these kinds of kit assays because of similarities in compound chemical structures and the often nonspecific ways in which the compounds are detected.

We report a case of three schizophrenia patients with quetiapine monotherapy who underwent urine drug screens in the context of a functional magnetic resonance imaging (fMRI) study. At the time of investigation, the patients did not take other drugs.

Using the Cobas Integra Methadone II testkit by Roche (1) , we found methadone-positive results that were unexpected, given the patients’ medical histories. The Cobas Integra Methadone II testkit contains an in vitro diagnosticum for semiquantitative and qualitative methadone detection in human urine, with a threshold of 300 ng/ml for methadone positivity. In a second step, we tested blood samples taken on the day after the last quetiapine intake to cross check with mass spectrometry (LC/MS/MS). Chromatographic peaks were confirmed and quantified by tandem mass spectrometry on a Micromass Quattro Ultima, using the mass transitions 384 >253 amu for quetiapine and 310 >265 amu for methadone as described previously (2) . Samples (190 microliter) were spiked with 10 microliter d3-methadone internal standard (1 microgram/ml in methanol, mass transition 268 >265) and extracted with 500 microliter –20°C cold acetonitrile. Hundred microliter supernatant were directly injected into the HPLC/MS/MS instrumentation. Chromatographic separation of the analytes was performed on a reverse-phase C18 column (Waters Acquity, 1.7 micrometer, 2.1×50 mm [www.waters.com]), with a mobile phase gradient from 50% acetonitrile and 50% 5 mM formic acid in water (buffer A) to 100% acetontrile (buffer B) over 2.2 over 1.8 minutes at a flow rate of 0.25 ml/minute. The level of quantification was 1 ng/ml for quetiapine and 2 ng/ml for methadone. Quetiapine metabolites were not quantified (determined). In contrast to the rechecked urinalyses (again methadone-positive), the quantitative assay, which was performed at the same time, did not reveal methadone positivity in the patients’ plasma.

In summary, in case of clinical doubt, positive drug screenings should always be verified by more specific quantitative tests before confronting patients with results. Given our findings, this could be of special interest in patients treated with quetiapine who show methadone-positive urine drug screens.